Performing Auto Sensor Calibration and Zero Adjustment
1. On the sensor calibration screen, press 2(CALIB).
A screen like the one shown below will appear after the first sampling operation starts.
First sampling operation
Real-time display of sampled values
2. After the sampled value stabilizes, hold down w for a few seconds.
This registers the first sampled valued and displays it on the screen. At this time, the
cursor will appear at the bottom of the display, indicating that a reference value can be
input.
3. Input a reference value for the first sample value and then press w.
A screen like the one shown below will appear after the second sampling operation starts
automatically.
Second sampling operation
4. After the sampled value stabilizes, hold down w for a few seconds.
This registers second sampled valued and displays it on the screen. At this time, the
cursor will appear at the bottom of the display, indicating that a reference value can be
input.
5. Input a reference value for the second sample value and then press w.
This returns to the sensor calibration screen.
E-CON4 calculates slope and
y-intercept values based on the two input reference values
and automatically configures settings. Automatically calculated values are displayed on
the sensor calibration screen.
Performing Sensor Zero Adjustment
1. On the sensor calibration screen, press 3(ZERO).
A screen like the one shown below will appear after sampling starts.