20021201
(3) Unit Name
Press 1(EDIT) to display a dialog box for inputting up to eight characters for the unit
name.
6. Press wand then input a memory number (1 to 99).
This saves the custom probe setup and returns to the Custom Probe List, which should
now contain the new custom probe setup you configured.
k Auto Calibrating a Custom Probe
Auto calibration automatically corrects the slope and intercept values of a custom probe
setup based on two actual samples.
Important!
Before performing the procedure below, you should prepare two conditions whose
measurement values are known.
When inputting reference value in step 4 of the procedure below, input the exact known
measurement value of the condition you will sample in step 5. When inputting reference
value in step 6 of the procedure below, input the exact known measurement value of the
condition you will sample in step 7.
uu
uu
u To auto calibrate a custom probe
1. Connect the calculator and EA-200, and connect the custom probe you want to auto
calibrate to CH1 of the EA-200.
2. Perform steps 1 through 4 of the procedure under “To configure a custom probe setup” on
page 4-1.
3. Press 2(CALIB).
This displays a dialog box like the one shown below.
4-2
Using a Custom Probe
4. Input a reference value for the first sample.
5. Press w.
This performs a sampling operation, and stores the value you input in step 4 and the
sampled value. Next, the dialog box for inputting a reference value reappears.
6. Input a reference value for the second sample.
7. Press w.
This performs a sampling operation, and stores the value you input in step 6 and the
sampled value.